Carbolite 3216CC Controller Manual de usuario

Busca en linea o descarga Manual de usuario para Panel De Control Carbolite 3216CC Controller. Carbolite 3216CC Controller User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
LMC20-GB v1.03
Operating Instructions
Temperature Controller
Type 3216CC
English
Vista de pagina 0
1 2 3 4 5 6 ... 23 24

Indice de contenidos

Pagina 1 - Type 3216CC

LMC20-GB – v1.03 Operating Instructions Temperature Controller  Type 3216CC English

Pagina 2

Operating 10 LMC20-GB 4.3 Customer ID. A furnace or oven identification number can be entered if required. This maybe used to identify one of many

Pagina 3 - Contents

Operating 11 LMC20-GB 5.2 Dwell Timer A dwell timer is used to control a process at a fixed temperature for a defined period. At the end of the ti

Pagina 4

Operating 12 LMC20-GB 5.4 Soft Start Timer The soft start timer is used to start a process at a reduced setpoint and power. The soft start setpoint

Pagina 5 - 1.1.1 3216CC - Controller

Operating 13 LMC20-GB 6 Running a Timer Run. This starts the timer. Hold. This stops the timer at the elapsed time. It will start again from the el

Pagina 6 - 2 Basic Operation

Operating 14 LMC20-GB Running Dwell Timer with Ramp Rate Set the ramp rate as outlined in section 4.1 Set the dwell time as outlined in section 5.2

Pagina 7 - 3 Quick Start Guide

Operating 15 LMC20-GB 6.2 Delayed Switch On Timer Operation Action Indication To RUN Timer Press and quickly release  +  Indicator – RUN = On S

Pagina 8 - Quick Start Guide

Operating 16 LMC20-GB Delay timer with ramp rate functions Timer Ended When ramp rate active Indicator – RUN = On - if ramp rate active Scrolling

Pagina 9 - 4 Setting up the controller

Operating 17 LMC20-GB Time Remaining The time remaining of any timer mode can be checked at any time while a timer is active. To view the time remai

Pagina 10 - 5 Timer

Options 18 LMC20-GB 7 Options Because options can be ordered in a variety of combinations and for a variety of purposes, exact instructions are not

Pagina 11 - LMC20-GB

Options 19 LMC20-GB 7.3 Comms Address Typically the comms address is set to 1, but this can be changed. In the case of RS485 and multiple instrumen

Pagina 12

2 LMC20-GB

Pagina 13 - 6 Running a Timer

Navigation 20 LMC20-GB 8 Navigation Diagram 8.1 3216CC Passcode = 9 LEVEL 1 ACCESS LEVEL 2PV LEV PVSP GOTO SP

Pagina 14

Controller Fault 21 LMC20-GB 9 Controller Fault 9.1 Fault Code Diagnostic Table Error Code Explanation Actions S.br Temperature sensor failure Rep

Pagina 15

Notes 22 LMC20-GB

Pagina 16

Notes 23 LMC20-GB

Pagina 17

Copyright © 2014 Carbolite Limited Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 62

Pagina 18 - 7 Options

3216CC Controller 3 MLC20-GB Contents 1 INTRODUCTION TO THE CONTROLLER AND MANUAL 5 1.1 USING THIS MANUAL 5 1.2 PID CONTROL 5 2 BASIC OPERATI

Pagina 20 - 8 Navigation Diagram

Basic Operation 5 LMC20-GB 1 Introduction to the Controller and Manual 1.1 Using This Manual This manual aims to explain how to set up and operate t

Pagina 21 - 9 Controller Fault

Introduction 6 LMC20-GB 2 Basic Operation 2.1 Controller Layout (Home Display) 2.2 Keys Page Key The Page key is used to acc

Pagina 22

Quick Start Guide 7 LMC20-GB 3 Quick Start Guide 3.1 Operation as a simple controller When switched on, the controller goes through a short test rou

Pagina 23

Quick Start Guide 8 LMC20-GB When level 2 operations have been completed, the supervisor must return to Level 1 either manually or by ‘power cycling’,

Pagina 24

Operating 9 LMC20-GB 4 Setting up the controller Before using the controller (or during its lifetime) certain parameters may have to be set, dependi

Comentarios a estos manuales

Sin comentarios